Concurrent Validity of Arterial Blood Oxygen Saturation Measurements: A Preliminary Analysis of an iPad Pulse Oximeter and Traditional Pulse Oximeter Using Bluetooth

in International Journal of Athletic Therapy and Training
Restricted access

Purchase article

USD  $24.95

Student 1 year subscription

USD  $74.00

1 year subscription

USD  $99.00

Student 2 year subscription

USD  $141.00

2 year subscription

USD  $188.00

Context:

Pulse oximetry has become mobile with the use of smartphone and Bluetooth wireless technology. This technology offers many benefits but has not been extensively studied. There is a need to further validate its clinimetric properties for health professionals to provide proper guidance to patients.

Objective:

This investigation assessed the concurrent validity of the iSpO2 pulse oximeter against a traditional pulse oximeter in measuring short-term resting blood oxygen saturation (SpO2) and pulse rate.

Design:

Observational study of reliability.

Setting:

University kinesiology laboratory.

Participants:

Thirty healthy, recre-ationally active adults (18 men, 12 women; mean age = 25.7 ± 5.46 years, mean height = 170.3cm ± 9.51, mean body mass = 76.4 kg ± 19.33).

Intervention:

Resting measurement of SpO2 and pulse rate using the iSpO2 pulse oximeter with the iPad Mini and a traditional pulse oximeter with Bluetooth.

Main Outcome Measure:

Resting SpO2 and pulse rate were concurrently measured over 5 min.

Results:

The concurrent validity between the iSpO2 and traditional pulse oximeter was moderate for measuring SpO2, intraclass correlation coeffcient (ICC)(3, 1) = .73, SEM = 0.70%, and good for pulse rate, ICC(3, 1) = .97, SEM = 1.74 beats per minute (bpm). The minimal detectable change at the 95% confidence interval for both instruments suggests that there may be 1.94% disagreement for SpO2 and 4.82 bpm disagreement between pulse oximetry methods. The 95% limits of agreement (LoA) for measuring SpO2 suggests that the iSpO2 and traditional pulse oximeters may vary -0.28 ± 1.98%, or approximately 2%. The 95% LoA for measuring pulse rate suggests that the iSpO2 and traditional pulse oximeter may vary 1.74 ± 4.98 bpm, potentially upward of 6 bpm. On the basis of the results of the LoA, it appears that there may be a slight systematic bias between the two devices, with the traditional pulse oximeter producing higher pulse rates than the iSpO2.

Conclusion:

The findings suggest that both instruments may be beneficial for indirect short-term measurements of resting SpO2 and pulse rate.