Validation of a New Handheld Device for Measuring Resting Metabolic Rate and Oxygen Consumption in Children

in International Journal of Sport Nutrition and Exercise Metabolism
Restricted access

Purchase article

USD $24.95

Student 1 year subscription

USD $87.00

1 year subscription

USD $116.00

Student 2 year subscription

USD $165.00

2 year subscription

USD $215.00

The purpose of this study was to assess the validity and reliability of the MedGem™ device to measure resting metabolic rate (RMR) in children. Subjects included 59 children (29 boys, 30 girls; mean age, 11.0 ± 0.2 y). Subjects were given 4 RMR tests during 1 test session, cconsisting of 2 Douglas bag and 2 MedGem tests, in random counterbalanced order. No significant differences were found between Douglas bag and MedGem systems for oxygen consumption (209 ± 5 and 213 ± 5 mL/min, respectively, P = 0.106, r = 0.911, mean ± standard deviation absolute difference 3.72 ± 17.40 mL/min) or RMR (1460 ± 39 and 1477 ± 35 kcal/d, P = 0.286, r = 0.909, mean ± standard deviation absolute difference 17.4 ± 124 kcal/d). Standard error of estimates for oxygen consumption and RMR were 17.4 mL/min and 124 kcal/d, respectively. In conclusion, these data indicate that the MedGem is a reliable and valid system for measuring oxygen consumption and RMR in children.

The authors are with the Dept of Health, Leisure, and Exercise Science, Appalachian State University, Boone, NC 28608.