The limited sample size in gait studies has hampered progress in the field. This challenge could be addressed through multicenter studies, thereby leveraging data sets from different laboratories. This study compared 3-dimensional lower-extremity running kinematics between the Biomechanics and Motor Control Laboratory, Federal University of ABC (Brazil), and the Running Injury Clinic, University of Calgary (Canada). Three-dimensional lower-extremity kinematics from 23 male runners were collected from each laboratory using comparable instrumentation and experimental procedures. The 3-dimensional hip, knee, and ankle angles were compared within and between centers using root-mean-square deviation. Two-sample t tests Statistical Parametric Mapping tested the hypothesis that the data from both laboratories were not different. The sagittal plane hip, knee, and ankle angles were similar between laboratories, while notable differences were observed for frontal (hip and ankle) and transverse (hip and knee) plane angles. The average interlaboratory root-mean-square deviation (2.6°) was lower than the intralaboratory root-mean-square deviation (Biomechanics and Motor Control = 4.8°, Running Injury Clinic = 5.6°), with the ankle transverse angle displaying the smallest, and the knee transverse angle displaying the largest variability. This study demonstrates the potential of combining gait kinematics data from different laboratories to increase sample size, but frontal and transverse plane data should be considered with caution.