Reliability and Validity of the VmaxST Portable Metabolic Analyzer

in Journal of Physical Activity and Health
Restricted access

Purchase article

USD $24.95

Student 1 year subscription

USD $115.00

1 year subscription

USD $153.00

Student 2 year subscription

USD $218.00

2 year subscription

USD $285.00

Background:

The reliability and validity of the SensorMedics VmaxST was tested.

Methods:

Thirty subjects (age = 24.5 ± 4.0 years, height = 174.8 ± 9.8 cm, weight = 70.3 ± 12.6 kg) performed treadmill exercise on three occasions, twice using the VmaxST and once using the SensorMedics 2900 system. Oxygen consumption (VO2; L/min) and heart rate (HR; beats/min) were measured continuously during three, 6- minute stages: 80 m/min, 0% grade; 94 m/min, 5% grade; and 160 m/min, 0% grade, and VO2max.

Results:

Reliability was high, and measurement error was low for VO2 (Rxx range = 0.97 - 0.99, CI = 0.94 - 1.00, SEM = 0.03 - 0.08 L/min) and HR (Rxx = 0.94 - 0.99, CI = 0.88 - 1.00, SEM = 1.8 - 3.2 beats/min). Validity was high for VO2 (Rxy range = 0.92 - 0.98, CI = 0.84 - 0.99, SEE = 0.08 - 0.21 L/min) and HR (Rxy = 0.97 - 0.99, CI = 0.94 - 1.00, SEE = 0.9 - 1.8 beats/min). Mean differences in VO2 between VmaxST and 2900 were small yet significant (P < 0.001).

Conclusions:

The VmaxST demonstrated excellent reliability and validity for measuring VO2 and HR over several exercise intensities. Small overestimates in VO2 by the VmaxST are countered by low measurement error.