Reliability of Inversion and Eversion Peak- and Average-Torque Measurements from the Biodex System 3 Dynamometer

Click name to view affiliation

Thomas W. Kaminski
Search for other papers by Thomas W. Kaminski in
Current site
Google Scholar
PubMed
Close
and
Geoff C. Dover
Search for other papers by Geoff C. Dover in
Current site
Google Scholar
PubMed
Close
Restricted access

Objective:

To determine the reliability of inversion and eversion concentric isokinetic-strength measurements from the Biodex System 3 isokinetic dynamometer.

Setting:

University biomechanics research laboratory.

Subjects:

Thirty-five volunteers free from any lower leg and ankle injuries within the preceding year.

Measurements:

Peak (PT) and average (AT) isokinetic torque at 30°/s and 120°/s for subtalar-joint inversion and eversion.

Results:

PT intraclass correlation coefficients (ICC2,1) ranged from .54 to .92. AT ICC2,1 ranged from .55 to .91. These ICCs were good to excellent for both PT and AT at each speed and motion tested, except for fair ICCs produced from right-foot-eversion measurements at 30°/s.

Conclusions:

Inversion and eversion subtalar-joint strength measurements from the Biodex System 3 isokinetic dynamometer are reliable.