Automated Exercise Blood Pressure Measurements in Children: A Preliminary Study

in Pediatric Exercise Science
Restricted access

Purchase article

USD  $24.95

Student 1 year subscription

USD  $68.00

1 year subscription

USD  $90.00

Student 2 year subscription

USD  $129.00

2 year subscription

USD  $168.00

This study compared blood pressure and heart rate measurements provided by the Quinton 410 automated exercise monitor with simultaneous auscultatory and electrocardiograph readings during rest, postural change, immediately after each treadmill exercise workload, and during active recovery in a sample of young children (mean age 7.9 yrs). The Quinton 410 provided highly accurate heart rates under all conditions (average mean difference <1.0 bpm). The Quinton systolic readings correlated well with and were similar to auscultation across conditions except for the initial treadmill workload. Slightly weaker relationships were observed between the Quinton and K4 diastolic comparisons. Compared with K4 auscultatory readings, the Quinton 410 provided slightly lower diastolic pressures across conditions (average mean difference = 3.1 mmHg). These findings provide preliminary evidence that for group comparisons with children, the Quinton 410 provides acceptable blood pressure estimates resulting from a variety of events, including exercise.

The authors are with the Medical College of Georgia, Augusta, GA 30912. Treiber, Strong, and Baranowski are with the Dept. of Pediatrics and the Georgia Institute for the Prevention of Human Disease and Accidents. McCaffrey is with the Dept. of Pediatrics. Davis is with the Office of Research Statistics and Computing.